Alaris Pump Pressure: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
No edit summary |
||
Line 18: | Line 18: | ||
# Text "IV " and choose IV Fluid Bolus | # Text "IV " and choose IV Fluid Bolus | ||
# Change the Container Volume and Dose to 900mL | # Change the Container Volume and Dose to 900mL | ||
# Change rate to | # Change rate to 400mL | ||
# Push Start and Click "Yes" to override | # Push Start and Click "Yes" to override ( *No longer necessary to change PSI* ) | ||
# Select previous screen and start infusion | # Select previous screen and start infusion | ||
[[Category:Setups]] | [[Category:Setups]] |
Latest revision as of 18:11, 27 October 2022
This page is a guide to overriding the pressure settings on the Alaris IV pump. Primarily used for the Sapien M3 setup.
Steps
- Press the white “Select Channel” button on the pump module.
- Select “Basic Infusion” on the brain.
- Change the rate to 200 ml/hr
- Change the VTBI to 900ml
- Press the options button on the brain at the bottom left of the display.
- Select “Pressure Limit”
- Select the “Selectable” option
- Using the up and buttons, change the pressure limit to 400 mmHg
Plum360 Pump Set Up
- Select Procedural from drug library
- Select pump A
- Text "IV " and choose IV Fluid Bolus
- Change the Container Volume and Dose to 900mL
- Change rate to 400mL
- Push Start and Click "Yes" to override ( *No longer necessary to change PSI* )
- Select previous screen and start infusion