Alaris Pump Pressure: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
No edit summary
 
Line 18: Line 18:
# Text "IV " and choose IV Fluid Bolus
# Text "IV " and choose IV Fluid Bolus
# Change the Container Volume and Dose to 900mL
# Change the Container Volume and Dose to 900mL
# Change rate to 200mL
# Change rate to 400mL
# Push Start and Click "Yes" to override  
# Push Start and Click "Yes" to override ( *No longer necessary to change PSI* )
# Select Settings and select pressure (Max 15 psi - this part is going to be trial by error)
# Select previous screen and start infusion
# Select previous screen and start infusion


[[Category:Setups]]
[[Category:Setups]]

Latest revision as of 18:11, 27 October 2022

This page is a guide to overriding the pressure settings on the Alaris IV pump. Primarily used for the Sapien M3 setup.

Steps

  1. Press the white “Select Channel” button on the pump module.
  2. Select “Basic Infusion” on the brain.
  3. Change the rate to 200 ml/hr
  4. Change the VTBI to 900ml
  5. Press the options button on the brain at the bottom left of the display.
  6. Select “Pressure Limit”
  7. Select the “Selectable” option
  8. Using the up and buttons, change the pressure limit to 400 mmHg


Plum360 Pump Set Up

  1. Select Procedural from drug library
  2. Select pump A
  3. Text "IV " and choose IV Fluid Bolus
  4. Change the Container Volume and Dose to 900mL
  5. Change rate to 400mL
  6. Push Start and Click "Yes" to override ( *No longer necessary to change PSI* )
  7. Select previous screen and start infusion