TMVI: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
No edit summary |
||
Line 25: | Line 25: | ||
*Micropuncture | *Micropuncture | ||
*8fr Pinnacle for Perclose | *8fr Pinnacle for Perclose | ||
8x35 Brite tip | *Perclose | ||
*8x35 Brite tip | |||
*150 J wire | *150 J wire | ||
*.'''032''' x 260 J-wire | *.'''032''' x 260 J-wire | ||
Line 35: | Line 32: | ||
*Amplatz Super Stiff 1cm tip | *Amplatz Super Stiff 1cm tip | ||
*8.5F SL1 | |||
*Baylis Needle | |||
* Baylis | |||
*Baylis cable | *Baylis cable | ||
*Grounding pad | *Grounding pad | ||
*8.5F Agilis | |||
*16F E sheath, always, regardless of valve size. | |||
*Naked 3 Port Manifold | |||
*6fr. Straight Pig | *6fr. Straight Pig | ||
*5fr. MPA2 | *5fr. MPA2 | ||
=== | |||
* | === SEPTAL DILATION === | ||
* | *12X4X75 MUSTANG '''for 23mm and 26mm Sapien OR''' 14mm Armada '''for 29mm Sapien''' | ||
*PRESTO INDEFLATOR | |||
===PACING=== | ===PACING=== | ||
*5F Josephson Pacing Cable | *5F Josephson Pacing Cable | ||
Line 54: | Line 51: | ||
*(2) Pacer Boxes | *(2) Pacer Boxes | ||
===FOR EDWARDS VALVE=== | ===FOR EDWARDS VALVE=== | ||
*Indeflator | *Indeflator (Second Indeflator on standby) | ||
*Crimper | *Crimper | ||
*E-sheath | *E-sheath (Second E-Sheath on standby) | ||
*Delivery system | *Delivery system (Second delivery on standby) | ||
*Valve | *Valve | ||
==Standby== | ==Standby== | ||
*9X4X75 MUSTANG | *9X4X75 MUSTANG | ||
Line 73: | Line 67: | ||
*EPINEPHRINE 10MCG/ML IN RED SYRINGE ON BACK TABLE | *EPINEPHRINE 10MCG/ML IN RED SYRINGE ON BACK TABLE | ||
*ROTAGLIDE 10ML (EDWARDS VALVE) | *ROTAGLIDE 10ML (EDWARDS VALVE) | ||
==PROCEDURE:== | ==PROCEDURE:== | ||
Line 104: | Line 80: | ||
*MP and SL1 removed. E Sheath advanced. | *MP and SL1 removed. E Sheath advanced. | ||
*Tourguide advanced and directed to mitral valve. | *Tourguide advanced and directed to mitral valve. | ||
*Pigtail and Safari2 inserted through | *Pigtail and Safari2 inserted through Agilis for valve crossing. | ||
*Safari2 wire inserted to ventricle. Simultaneous LA/LV pressure recorded. | *Safari2 wire inserted to ventricle. Simultaneous LA/LV pressure recorded. | ||
*Pigtail and | *Pigtail and Agilis removed. | ||
*12mm/14mm balloon advanced for septostomy and possible BAV | *12mm/14mm balloon advanced for septostomy and possible BAV | ||
*Balloon removed, valve delivery system advanced. | *Balloon removed, valve delivery system advanced. | ||
*When valve is in place, temp pacer inserted and tested. | *When valve is in place, temp pacer inserted and tested. | ||
*Valve deployed. Septum evaluated with echo for shunting. | *Valve deployed. Septum evaluated with echo for shunting. | ||
*Delivery removed and SL1 | *Delivery removed and SL1 re-advanced. | ||
*Pigtail advanced to LV, wire removed and simultaneous LA/LV pressures recorded. | *Pigtail advanced to LV, wire removed and simultaneous LA/LV pressures recorded. | ||
*Pigtail and SL1 removed. | *Pigtail and SL1 removed. | ||
Line 118: | Line 94: | ||
*8F vein pulled and manual pressure applied | *8F vein pulled and manual pressure applied | ||
*Dermabond access site, no dressings needed. | *Dermabond access site, no dressings needed. | ||
== STANDBY EMERGENCY ECMO SUPPLIES == | |||
* 16F ARTERIAL CANNULA | |||
*21F VENOUS CANNULA | |||
*(2) PERCLOSE | |||
*ECMO INSTRUMENT TRAY | |||
*AMPLATZ SUPER STIFF | |||
==STANDBY PERICARDIOCENTESIS SUPPLIES== | |||
*EYE DRAPE | |||
*CHIBA NEEDLE 10CM 18G | |||
*SKATER INTRODUCER SET | |||
*BARD LOCKING PIGTAIL 6F OR 8F | |||
*(3) 60ML SYRINGES | |||
*STOPCOCK | |||
*AMPLATZ SUPER STIFF 180CM | |||
<br> | <br> |
Revision as of 15:29, 2 September 2022
Warning: This page needs revision!
Protip: Need Alcohol in lab for this procedure!
TransCatheter Mitral Valve Repair/Replacement
Per Rosalie:
TAVR/TMVI
- Edwards Sapien 3 Ultra (20, 23 & 26mm) and Sapien 3 (29mm) – scan the valve barcode (box) in McKesson for documentation and charges. Bundle includes valve, delivery system, crimper, e sheath intro set, balloon cath & locking syringe. With the new inventory system (POU) space is limited in entering item full description. If pre dilatation Edwards balloon is necessary free text the balloon used in procedure notes (McKesson).
- Medtronic Core Valve – all items are charge separately. Scan all barcodes (delivery system, loading system and valve).
MITRAL CLIP (Mitral/Tricuspid):
- Use the barcode on Mckesson screen in Lab 1 – Mitraclip (Abbott) bundle consists of (1) Steerable Guide Catheter, (2) NTR Clip Delivery System & (1) XTR Clip Delivery System.
GENERAL ANESTHESIA AND TEE IMAGING, RADIAL ART LINE BY ANESTHESIA
ACCESS 2 FEMORAL VEINS (TAVR AND PACER)
PRE PROCEDURE: TYPE AND CROSS, ASPIRIN GIVEN
Open:
SHEATHS:
- Micropuncture
- 8fr Pinnacle for Perclose
- Perclose
- 8x35 Brite tip
- 150 J wire
- .032 x 260 J-wire
- Safari 2
- Amplatz Super Stiff 1cm tip
- 8.5F SL1
- Baylis Needle
- Baylis cable
- Grounding pad
- 8.5F Agilis
- 16F E sheath, always, regardless of valve size.
- Naked 3 Port Manifold
- 6fr. Straight Pig
- 5fr. MPA2
SEPTAL DILATION
- 12X4X75 MUSTANG for 23mm and 26mm Sapien OR 14mm Armada for 29mm Sapien
- PRESTO INDEFLATOR
PACING
- 5F Josephson Pacing Cable
- EP Quad Cable
- Remington Pacing Cable
- (2) Pacer Boxes
FOR EDWARDS VALVE
- Indeflator (Second Indeflator on standby)
- Crimper
- E-sheath (Second E-Sheath on standby)
- Delivery system (Second delivery on standby)
- Valve
Standby
- 9X4X75 MUSTANG
- 14, 16 XXL Balloons
- TORQUEVUE 45DEGREE DELIVERY SYSTEM (9F OR 10F)
- 10MM, 14MM, 18MM SEPTAL OCCLUDER
MEDICATIONS
- BUPIVACAINE
- EPINEPHRINE 10MCG/ML IN RED SYRINGE ON BACK TABLE
- ROTAGLIDE 10ML (EDWARDS VALVE)
PROCEDURE:
- Patient prepped from navel to knees in sterile fashion.
- Access obtained with u/s
- Dilate with 8F dilator
- Preclose
- SL1 inserted over .032 260cm J ( Don't use wire that comes with SL1 - toss off table)
- Baylis needle inserted through SL1 and atrial septum puncture performed.
- Needle removed, LA pressure recorded. .032 wire advanced to pulmonary vein
- MP advanced to LUPV, .032 wire removed and super stiff advanced
- MP and SL1 removed. E Sheath advanced.
- Tourguide advanced and directed to mitral valve.
- Pigtail and Safari2 inserted through Agilis for valve crossing.
- Safari2 wire inserted to ventricle. Simultaneous LA/LV pressure recorded.
- Pigtail and Agilis removed.
- 12mm/14mm balloon advanced for septostomy and possible BAV
- Balloon removed, valve delivery system advanced.
- When valve is in place, temp pacer inserted and tested.
- Valve deployed. Septum evaluated with echo for shunting.
- Delivery removed and SL1 re-advanced.
- Pigtail advanced to LV, wire removed and simultaneous LA/LV pressures recorded.
- Pigtail and SL1 removed.
- Pacer removed.
- Preclosure completed and stitch to RFV
- 8F vein pulled and manual pressure applied
- Dermabond access site, no dressings needed.
STANDBY EMERGENCY ECMO SUPPLIES
- 16F ARTERIAL CANNULA
- 21F VENOUS CANNULA
- (2) PERCLOSE
- ECMO INSTRUMENT TRAY
- AMPLATZ SUPER STIFF
STANDBY PERICARDIOCENTESIS SUPPLIES
- EYE DRAPE
- CHIBA NEEDLE 10CM 18G
- SKATER INTRODUCER SET
- BARD LOCKING PIGTAIL 6F OR 8F
- (3) 60ML SYRINGES
- STOPCOCK
- AMPLATZ SUPER STIFF 180CM
APPROVED: MD initials MM/YY