TMVI - Encircle Sapien M3: Difference between revisions
From Bay Area Structural Heart Wiki
(→Misc:) |
(→Misc:) |
||
Line 2: | Line 2: | ||
{{Protip|Need Alcohol in lab for this procedure!}} | {{Protip|Need Alcohol in lab for this procedure!}} | ||
{{Warn|<BR>THIS IS A RESEARCH PROCEDURE. ALL EDWARDS EQUIMENT WILL BE SUPPLIED BY THE RESEARCH TEAM.<BR> DO NOT THROW AWAY ANY BOXES.}} | {{Warn|<BR>THIS IS A RESEARCH PROCEDURE. ALL EDWARDS EQUIMENT WILL BE SUPPLIED BY THE RESEARCH TEAM.<BR> DO NOT THROW AWAY ANY BOXES. DOCUMENTATION NOTES IN RESEARCH HELPER IN MCKESSON. }} | ||
{{Warn|If broadcasting case:<BR>DO NOT PUT PATIENT INFO INTO MCKESSON OR XRAY. PHI CANNOT BE STREAMED. <BR>CASE INFO WILLL NEED TO BE MERGED AFTER THE PROCEDURE. PLEASE DO NOT SUBMIT CHARGES, ROSALIE WILL DO THEM. }} | {{Warn|If broadcasting case:<BR>DO NOT PUT PATIENT INFO INTO MCKESSON OR XRAY. PHI CANNOT BE STREAMED. <BR>CASE INFO WILLL NEED TO BE MERGED AFTER THE PROCEDURE. PLEASE DO NOT SUBMIT CHARGES, ROSALIE WILL DO THEM. }} | ||
Line 49: | Line 49: | ||
*5F MPA1 125 | *5F MPA1 125 | ||
===Pacing:=== | ===Pacing:=== | ||
*5F Josephson Pacing Cable | Unipolar pacing setup with Bipolar on standby, will still get LFV access | ||
*EP Quad Cable | |||
*Remington Pacing Cable | * Alligator Pacing Cable (open) | ||
*5F Josephson Pacing Cable (Standby) | |||
*EP Quad Cable (Standby) | |||
*Remington Pacing Cable (Standby) | |||
===Misc:=== | ===Misc:=== | ||
*Extra Sterile Bowl (3- 0.9% SALINE, 1-HEP SALINE, 1-15% CONTRAST MIX) | *Extra Sterile Bowl (3- 0.9% SALINE, 1-HEP SALINE, 1-15% CONTRAST MIX) | ||
Line 58: | Line 62: | ||
*(3) High Pressure stopcock | *(3) High Pressure stopcock | ||
*(1) Heparin tubing (updated with new device) | *(1) Heparin tubing (updated with new device) | ||
*( | *(2) 72” pressure tubing (updated with new device) | ||
*Extra manifold (For a gradient at the end) | *Extra manifold (For a gradient at the end) | ||
*Perclose | *(1) Perclose | ||
*0-Prolene | *0-Prolene | ||
===Baylis:=== | ===Baylis:=== |
Revision as of 17:39, 24 March 2022
Protip: Need Alcohol in lab for this procedure!
Warning:
THIS IS A RESEARCH PROCEDURE. ALL EDWARDS EQUIMENT WILL BE SUPPLIED BY THE RESEARCH TEAM.
DO NOT THROW AWAY ANY BOXES. DOCUMENTATION NOTES IN RESEARCH HELPER IN MCKESSON.
Warning: If broadcasting case:
DO NOT PUT PATIENT INFO INTO MCKESSON OR XRAY. PHI CANNOT BE STREAMED.
CASE INFO WILLL NEED TO BE MERGED AFTER THE PROCEDURE. PLEASE DO NOT SUBMIT CHARGES, ROSALIE WILL DO THEM.
Sapien M3 | |||||
---|---|---|---|---|---|
Anesthesia | Imaging | Access | Pre-Procedure | ||
|
|
|
|
Open:
Sheaths:
- Micropuncture set
- 8f Pinnacle
- 8fr x 35cm Brite-Tip
- 8.5fr SL1 transseptal sheath
- 22F coons dilator
Wires:
- (1) .035 150cm J wire
- (1) .035 260cm J wire
- (1) .032 260cm J wire
- (1) Safari Wire
- (1) Amplatz SS 260cm 1cm Tip
Catheters:
- (1) 6F Straight Pigtail
- 5F MPA1 125
Pacing:
Unipolar pacing setup with Bipolar on standby, will still get LFV access
- Alligator Pacing Cable (open)
- 5F Josephson Pacing Cable (Standby)
- EP Quad Cable (Standby)
- Remington Pacing Cable (Standby)
Misc:
- Extra Sterile Bowl (3- 0.9% SALINE, 1-HEP SALINE, 1-15% CONTRAST MIX)
- (3) 60ml Syringe
- (3) 20ml Syringe
- (3) High Pressure stopcock
- (1) Heparin tubing (updated with new device)
- (2) 72” pressure tubing (updated with new device)
- Extra manifold (For a gradient at the end)
- (1) Perclose
- 0-Prolene
Baylis:
- Baylis needle
- Baylis cable
Standby:
- Baylis ProTrack wire
- 6F ballon wedge
- 2530 Edwards indeflator
- 14X4 Armada Balloon
- PRESTO INDEFLATOR
- TORQUEVUE 45 DEGREE DELIVERY SYSTEM (9F OR 10F)
- 10MM, 14MM, 18MM SEPTAL OCCLUDER
Nurse Notes:
- ACIST- 100% CONTRAST
- 1L Hep Saline on Alaris IV pump set at: Pressure: 400 mmHg Rate: 400ml/hr - Click this link to see pressure adjustment process
- Grounding pad
- Baylis machine
- (2) Pacer Boxes
- Patient prepped from navel to knees in sterile fashion.
- (1) Alaris pump (updated with new device)
MEDICATIONS
- Bupivicaine/Lidocaine Mixed 10mL each
- Epinephrine 10mcg/mL in red syringe on back table
- Rotaglide 10mL for sheath before valve is inserted
PROCEDURE
- Access obtained with u/s
- Need 20 and 60 cc flush on front table for the delivery system.
- 8F 35cm LFV for transcutaneous pacing
- RFV
- Dilate with 8F
- Preclose
- SL1 inserted over .032x260 wire
- Baylis needle inserted through SL1 and atrial septum puncture performed.
- Needle removed
- 5F MPA advanced over .032 wire
- 1cm Amplatz Super Stiff advanced and MPA and SL1 removed
- 23fr. Delivery sheath(GS) - Used for Doc and valve advanced
- Pigtail advanced across the mitral valve and LVgram @ 15/45 performed
- Mitral annulus denoted on live monitor screen in A plane
- DDS prepped
- 1L Hep. Saline (2U/ml) bag with pressure bag spiked with heparin tubing, flushed, and attached to pump
- Alaris settings 400ml/hr occlusion pressure limit 400mm Hg
- Pressure tubing attached to stopcock and DDS
- Swiftlink cover installed over pressure tubing and secured ~18” from off-table end
- After hookup and flush, Swiftlink cover can be slid over heparin tubing for additional slack on the table.
- Dock Delivery System (DDS) advanced through GS
- Dock deployed in a timely fashion ;)
- Temp pacer inserted and tested.
- DDS removed and Pigtail inserted over exchange J for valve crossing
- J wire exchanged for Safari 2
- Pigtail and GS removed
- Balloon removed
- Valve delivery system advanced
- Valve deployed. Septum evaluated with echo for shunting and PVL.
- remind MD to lower FIO2*** Delivery system removed and MPA inserted for LA pressure measurement.
- (RHC performed with 6F swan to evaluate ASD)
- Safari retained across septum until shunt is evaluated
- All equipment removed
- Preclosure closed and stitch to RFV, manual pressure LFV
- Woggle vs Dermabond depending on Protamine
STANDBY EMERGENCY ECMO SUPPLIES
- 16F Arterial Cannula
- 21F Venous Cannula
- (2) Perclose
- ECMO Instrument Tray
- Amplatz SS 7cm tip
STANDBY PERICARDIOCENTESIS SUPPLIES
- EYE DRAPE
- CHIBA NEEDLE 10CM 18G
- SKATER INTRODUCER SET
- BARD LOCKING PIGTAIL 6F OR 8F
- (3) 60ML SYRINGES
- STOPCOCK
- AMPLATZ SUPER STIFF 180CM
STANDBY PARAVALVULAR LEAK REPAIR SUPPLIES
SEE MITRAL PARAVALVULAR LEAK REPAIR
- AMPLATZER VASCULAR PLUG 2 SIZE 8-18 AVAILABLE
APPROVED: MD initials MM/YY