Septostomy: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
Line 4: Line 4:


An atrial Septostomy is performed in the setting of ECMO when there is a concern that Left Atrial pressures are high due to the ECMO flow. Generally done in the setting of ECMO. It can be a stand alone procedure if not performed when ECMO is initiated.
An atrial Septostomy is performed in the setting of ECMO when there is a concern that Left Atrial pressures are high due to the ECMO flow. Generally done in the setting of ECMO. It can be a stand alone procedure if not performed when ECMO is initiated.
ECHO guided transeptal puncture.
<br />


===Equipment:===
===Equipment:===
====Open====
====Open====
*Amplatz Super Stiff 1cm
*Amplatz Super Stiff 1cm
*.032 x 260J
*.032 x 260J
Line 22: Line 16:
*0-Prolene
*0-Prolene


<br />


====Standby====
====Standby====
*Amplatz Extra Stiff
*Amplatz Extra Stiff


===Baylis Setup:===
===Baylis Setup:===
 
====Open:====
====Open====
 
*Baylis generator
*Baylis generator
*Baylis Large curve needle
*Baylis Large curve needle
*Baylis cable
*Baylis cable
*Grounding Pad
*Grounding Pad
<br />


===Balloon size determined by MD.===  
===Balloon size determined by MD.===  
'''<nowiki>*</nowiki>Sheath size determined by balloon. Balloon volume listed on Structural Heart Cart, and chart above. '''
*Sheath size determined by balloon. Balloon volume listed on Structural Heart Cart, and chart above.
 
*Have a 22 Tyshak nucleus on standby for Spies and Daniels  
==== *Have a 22 Tyshak nucleus on standby for Spies and Daniels ====
<br />
<br />
<hr />
<hr />
Approved: MD initials MM/YY
Approved: MD initials MM/YY
[[Category:Procedures]]
[[Category:Procedures]]

Revision as of 21:29, 6 December 2021

Z-Med Inflation Chart
Other Balloons Inflation Chart

An atrial Septostomy is performed in the setting of ECMO when there is a concern that Left Atrial pressures are high due to the ECMO flow. Generally done in the setting of ECMO. It can be a stand alone procedure if not performed when ECMO is initiated.

Equipment:

Open

  • Amplatz Super Stiff 1cm
  • .032 x 260J
  • 5fr. MPA2
  • 8.5fr. SL1
  • Manifold to measure LA pressure
  • Sheath based on balloon size (See chart)
  • Presto indeflator
  • 0-Prolene


Standby

  • Amplatz Extra Stiff

Baylis Setup:

Open:

  • Baylis generator
  • Baylis Large curve needle
  • Baylis cable
  • Grounding Pad

Balloon size determined by MD.

  • Sheath size determined by balloon. Balloon volume listed on Structural Heart Cart, and chart above.
  • Have a 22 Tyshak nucleus on standby for Spies and Daniels



Approved: MD initials MM/YY