Double Pump Valve Cracker: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
No edit summary |
||
Line 8: | Line 8: | ||
*4-way stopcock | *4-way stopcock | ||
*True Balloon | *True Balloon | ||
** | **Fill big indeflator, 10ml in small indeflator | ||
==Procedure== | ==Procedure== |
Revision as of 19:04, 28 October 2021
This setup is used to over dilate an existing valve in preparation for installing a new trans catheter valve.
Equipment
- Atrion 38 indeflator
- Atrion 25 indeflator
- 4-way stopcock
- True Balloon
- Fill big indeflator, 10ml in small indeflator
Procedure
- Use 30/170 TAVR mix
- De-air and fill both indeflators
- Prep balloon similar to Edwards Delivery system prep
- Attach both indeflators and open stopcock to be open all ways
- Lock smaller indeflator
- Leave larger indeflator unlocked