Mitraclip PLUG: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
|||
(8 intermediate revisions by 3 users not shown) | |||
Line 4: | Line 4: | ||
*260cm J Wire | *260cm J Wire | ||
*Amplatz ES 260 | |||
*20F Dryseal | *20F Dryseal | ||
*Agilis | *Agilis | ||
*5F 110cm Shuttle Sheath (dependent on plug size) | *5F 110cm Shuttle Sheath (dependent on plug size) | ||
*5F 10cm Pinnacle Sheath | *5F 10cm Pinnacle Sheath (for shuttle sheath not access) | ||
*Straight glidewire (standard) | *Straight glidewire (standard) | ||
*180* torqvue (French size not important here, only using for the cable.)<br /> | *.032 x 260 J | ||
*MPA 125cm | |||
*180* torqvue (French size not important here, only using for the cable.) | |||
{{Note|The deployment cable for the torqvue can be daisy-chained for more length!}} | |||
*0-Ethibond | |||
*Naked 3-Port manifold for pressures only <br /> | |||
===Procedure for placing a plug between 2 | ===Procedure for placing a plug between 2 Mitraclips in the same setting:=== | ||
*8F sheath and 260cm J inserted through guide delivery system. | *8F sheath and 260cm J inserted through guide delivery system. | ||
Line 19: | Line 25: | ||
*Glidewire used to cross between plugs and removed with shuttle dilator. | *Glidewire used to cross between plugs and removed with shuttle dilator. | ||
*5F sheath inserted into shuttle in lieu of using torqvue delivery sideport | *5F sheath inserted into shuttle in lieu of using torqvue delivery sideport | ||
**'''Transition between shuttle and torqvue insertion isn't flush so occluder starts to deploy if sheath not | **'''Transition between shuttle and torqvue insertion isn't flush so occluder starts to deploy if sheath not used''' | ||
== CARDIOFORM == | ==CARDIOFORM== | ||
=== Equipment: === | ===Equipment:=== | ||
* Angled Glide | '''14fr. MTS = Mullins''' | ||
* Amplatz ES | |||
* Amplatz SS | *Angled Glide | ||
* Rosen Wire | *Amplatz ES | ||
* 5F MP | *Amplatz SS | ||
* 7F Shuttle - dilator used inside Cardioform sheath | *Rosen Wire | ||
* Flexcath- ***flexcath $150, Agilis $900*** | *5F MP | ||
* 20F Dryseal | *7F Shuttle - dilator used inside Cardioform sheath | ||
* 20mm Snare- snare dilator to angle toward valve | *Flexcath- ***flexcath $150, Agilis $900*** | ||
*20F Dryseal | |||
*20mm Snare- snare dilator to angle toward valve | |||
<br /> | <br /> | ||
==== IF Arterial Access needed for Body Floss ==== | ====IF Arterial Access needed for Body Floss==== | ||
*Micropuncture | |||
*6F Sheath | |||
*Guide | |||
*Watchdog | |||
*Stopcock | |||
*JAG wire | |||
*Flexcath used for direction but has to be removed because it's too long for cardioform. | |||
[[Category:Procedures]] | [[Category:Procedures]] |
Latest revision as of 20:26, 21 April 2021
ADOII
Equipment:
- 260cm J Wire
- Amplatz ES 260
- 20F Dryseal
- Agilis
- 5F 110cm Shuttle Sheath (dependent on plug size)
- 5F 10cm Pinnacle Sheath (for shuttle sheath not access)
- Straight glidewire (standard)
- .032 x 260 J
- MPA 125cm
- 180* torqvue (French size not important here, only using for the cable.)
Note: The deployment cable for the torqvue can be daisy-chained for more length!
- 0-Ethibond
- Naked 3-Port manifold for pressures only
Procedure for placing a plug between 2 Mitraclips in the same setting:
- 8F sheath and 260cm J inserted through guide delivery system.
- Guide delivery system removed, 20F dryseal inserted
- Agilis inserted
- Preload Shuttle sheath with glidewire and advance through agilis
- Glidewire used to cross between plugs and removed with shuttle dilator.
- 5F sheath inserted into shuttle in lieu of using torqvue delivery sideport
- Transition between shuttle and torqvue insertion isn't flush so occluder starts to deploy if sheath not used
CARDIOFORM
Equipment:
14fr. MTS = Mullins
- Angled Glide
- Amplatz ES
- Amplatz SS
- Rosen Wire
- 5F MP
- 7F Shuttle - dilator used inside Cardioform sheath
- Flexcath- ***flexcath $150, Agilis $900***
- 20F Dryseal
- 20mm Snare- snare dilator to angle toward valve
IF Arterial Access needed for Body Floss
- Micropuncture
- 6F Sheath
- Guide
- Watchdog
- Stopcock
- JAG wire
- Flexcath used for direction but has to be removed because it's too long for cardioform.