Mitraclip PLUG: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
 
(11 intermediate revisions by 3 users not shown)
Line 1: Line 1:
== ADOII ==
==ADOII==


=== Equipment: ===
===Equipment:===


*260cm J Wire
*Amplatz ES 260
*20F Dryseal
*Agilis
*5F 110cm Shuttle Sheath (dependent on plug size)
*5F 10cm Pinnacle Sheath (for shuttle sheath not access)
*Straight glidewire (standard)
*.032 x 260 J
*MPA 125cm
*180* torqvue (French size not important here, only using for the cable.)
{{Note|The deployment cable for the torqvue can be daisy-chained for more length!}}
*0-Ethibond 
*Naked 3-Port manifold for pressures only <br />


===Procedure for placing a plug between 2 Mitraclips in the same setting:===
*8F sheath and 260cm J inserted through guide delivery system.
*Guide delivery system removed, 20F dryseal inserted
*Agilis inserted
*Preload Shuttle sheath with glidewire and advance through agilis
*Glidewire used to cross between plugs and removed with shuttle dilator.
*5F sheath inserted into shuttle in lieu of using torqvue delivery sideport
**'''Transition between shuttle and torqvue insertion isn't flush so occluder starts to deploy if sheath not used'''
==CARDIOFORM==
===Equipment:===
'''14fr. MTS = Mullins'''
*Angled Glide
*Amplatz ES
*Amplatz SS
*Rosen Wire
*5F MP
*7F Shuttle - dilator used inside Cardioform sheath
*Flexcath- ***flexcath $150, Agilis $900***
*20F Dryseal
*20mm Snare- snare dilator to angle toward valve


When placing a plug between 2 Mitraclip in the same setting:
<br />
<br />
====IF Arterial Access needed for Body Floss====
*Micropuncture
*6F Sheath
*Guide
*Watchdog
*Stopcock
*JAG wire
*Flexcath used for direction but has to be removed because it's too long for cardioform.
[[Category:Procedures]]

Latest revision as of 20:26, 21 April 2021

ADOII

Equipment:

  • 260cm J Wire
  • Amplatz ES 260
  • 20F Dryseal
  • Agilis
  • 5F 110cm Shuttle Sheath (dependent on plug size)
  • 5F 10cm Pinnacle Sheath (for shuttle sheath not access)
  • Straight glidewire (standard)
  • .032 x 260 J
  • MPA 125cm
  • 180* torqvue (French size not important here, only using for the cable.)

Note Note: The deployment cable for the torqvue can be daisy-chained for more length!

  • 0-Ethibond
  • Naked 3-Port manifold for pressures only

Procedure for placing a plug between 2 Mitraclips in the same setting:

  • 8F sheath and 260cm J inserted through guide delivery system.
  • Guide delivery system removed, 20F dryseal inserted
  • Agilis inserted
  • Preload Shuttle sheath with glidewire and advance through agilis
  • Glidewire used to cross between plugs and removed with shuttle dilator.
  • 5F sheath inserted into shuttle in lieu of using torqvue delivery sideport
    • Transition between shuttle and torqvue insertion isn't flush so occluder starts to deploy if sheath not used


CARDIOFORM

Equipment:

14fr. MTS = Mullins

  • Angled Glide
  • Amplatz ES
  • Amplatz SS
  • Rosen Wire
  • 5F MP
  • 7F Shuttle - dilator used inside Cardioform sheath
  • Flexcath- ***flexcath $150, Agilis $900***
  • 20F Dryseal
  • 20mm Snare- snare dilator to angle toward valve


IF Arterial Access needed for Body Floss

  • Micropuncture
  • 6F Sheath
  • Guide
  • Watchdog
  • Stopcock
  • JAG wire
  • Flexcath used for direction but has to be removed because it's too long for cardioform.