Cardiomems: Difference between revisions

From Bay Area Structural Heart Wiki
(Created page with "The Cardiomems device is used to passively monitor Pulmonary Artery pressures. ==Dr. Herr== ===Open=== *US probe cover *Micropuncture *75cm or 180 cm Amplatz Super Stiff *7f...")
 
 
(11 intermediate revisions by 2 users not shown)
Line 1: Line 1:
The Cardiomems device is used to passively monitor Pulmonary Artery pressures.
The Cardiomems device is used to passively monitor Pulmonary Artery pressures.
====Procedure type in McKesson: (33289) IMPL WRLS PA PRESSURE SENSOR====
*Under Diagnostic heart cath.
===Reps===
*Dan Kivelson 925 787-5551
*Robert Tubbs 669 232-4527




==Dr. Herr==
==Dr. Herr==
===Open===
===Open===
*US probe cover
*US probe cover
*Micropuncture
*Micropuncture
Line 11: Line 20:
*7fr. TD swan
*7fr. TD swan
*.025 wire
*.025 wire
*Steelcore
*.018 x 300 Steelcore
*2-port Pressure manifold
*2-port Pressure manifold
*~20cc contrast
*~20cc contrast
*Glowtape (Usually behind room 6)
{{Warn| The glowtape expired, so it has been removed from the sterile packaging, and lives on the shelf in Room 6 above the CardioMems devices. Sterility not needed, Place behind the back as usual.}}


===Standby===
===Standby===
*Cardiomems Device
*Cardiomems Device




===Procedure===
===Procedure===
{{Note|No need for contrast manifold, Dr. Herr prefers to inject with a syringe}}
#Place Glowtape under patient, slightly left of spine, mid chest
#Right groin access with micropuncture and ultrasound
#Amplatz wire inserted in micropuncture sheath
#7 and 9 French dilators
#12 French Fast Cath
#Swan inserted for full Right Heart Cath (.025 wire used if needed to get to PA)
#Dr. Herr prefers Left PA for the device
#Pulmonary Arteriogram for anatomy and placement planning (Glowtape provides reference point)
#Open device, flush wire lumen, and whisk in saline for 10-30 seconds
#Insert device and advance to PA
#Device is deployed, delivery system removed, Steelcore pulled back to common PA
#Swan advanced over steelcore to right PA for calibration numbers
#Cardiomems calibrated
#Sheath pulled manual pressure only.
[[Category:Procedures]]

Latest revision as of 15:09, 16 March 2021

The Cardiomems device is used to passively monitor Pulmonary Artery pressures.


Procedure type in McKesson: (33289) IMPL WRLS PA PRESSURE SENSOR

  • Under Diagnostic heart cath.

Reps

  • Dan Kivelson 925 787-5551
  • Robert Tubbs 669 232-4527


Dr. Herr

Open

  • US probe cover
  • Micropuncture
  • 75cm or 180 cm Amplatz Super Stiff
  • 7fr. And 9fr. dilators
  • 12fr. FastCath sheath
  • 7fr. TD swan
  • .025 wire
  • .018 x 300 Steelcore
  • 2-port Pressure manifold
  • ~20cc contrast
  • Glowtape (Usually behind room 6)

Note Warning: The glowtape expired, so it has been removed from the sterile packaging, and lives on the shelf in Room 6 above the CardioMems devices. Sterility not needed, Place behind the back as usual.

Standby

  • Cardiomems Device


Procedure

Note Note: No need for contrast manifold, Dr. Herr prefers to inject with a syringe

  1. Place Glowtape under patient, slightly left of spine, mid chest
  2. Right groin access with micropuncture and ultrasound
  3. Amplatz wire inserted in micropuncture sheath
  4. 7 and 9 French dilators
  5. 12 French Fast Cath
  6. Swan inserted for full Right Heart Cath (.025 wire used if needed to get to PA)
  7. Dr. Herr prefers Left PA for the device
  8. Pulmonary Arteriogram for anatomy and placement planning (Glowtape provides reference point)
  9. Open device, flush wire lumen, and whisk in saline for 10-30 seconds
  10. Insert device and advance to PA
  11. Device is deployed, delivery system removed, Steelcore pulled back to common PA
  12. Swan advanced over steelcore to right PA for calibration numbers
  13. Cardiomems calibrated
  14. Sheath pulled manual pressure only.