====Procedure type in McKesson: (33289) IMPL WRLS PA PRESSURE SENSOR==== | ====Procedure type in McKesson: (33289) IMPL WRLS PA PRESSURE SENSOR==== | ||
*Under Diagnostic heart cath. | *Under Diagnostic heart cath. | ||
===Reps=== | |||
*Dan Kivelson 925 787-5551 | |||
*Robert Tubbs 669 232-4527 | |||
Line 20: | Line 24: | ||
*~20cc contrast | *~20cc contrast | ||
*Glowtape (Usually behind room 6) | *Glowtape (Usually behind room 6) | ||
{{Warn| The glowtape expired, so it has been removed from the sterile packaging, and lives on the shelf in Room 6 above the CardioMems devices. Sterility not needed, Place behind the back as usual.}} | |||
===Standby=== | ===Standby=== | ||
Line 28: | Line 33: | ||
===Procedure=== | ===Procedure=== | ||
{{Note|No need for contrast manifold, Dr. Herr prefers to inject with a syringe}} | |||
#Place Glowtape under patient, slightly left of spine, mid chest | #Place Glowtape under patient, slightly left of spine, mid chest | ||
#Right groin access with micropuncture and ultrasound | #Right groin access with micropuncture and ultrasound | ||
#Amplatz wire inserted in micropuncture sheath | #Amplatz wire inserted in micropuncture sheath |
Latest revision as of 15:09, 16 March 2021
The Cardiomems device is used to passively monitor Pulmonary Artery pressures.
Procedure type in McKesson: (33289) IMPL WRLS PA PRESSURE SENSOR
- Under Diagnostic heart cath.
Reps
- Dan Kivelson 925 787-5551
- Robert Tubbs 669 232-4527
Dr. Herr
Open
- US probe cover
- Micropuncture
- 75cm or 180 cm Amplatz Super Stiff
- 7fr. And 9fr. dilators
- 12fr. FastCath sheath
- 7fr. TD swan
- .025 wire
- .018 x 300 Steelcore
- 2-port Pressure manifold
- ~20cc contrast
- Glowtape (Usually behind room 6)
Warning: The glowtape expired, so it has been removed from the sterile packaging, and lives on the shelf in Room 6 above the CardioMems devices. Sterility not needed, Place behind the back as usual.
Standby
- Cardiomems Device
Procedure
Note: No need for contrast manifold, Dr. Herr prefers to inject with a syringe
- Place Glowtape under patient, slightly left of spine, mid chest
- Right groin access with micropuncture and ultrasound
- Amplatz wire inserted in micropuncture sheath
- 7 and 9 French dilators
- 12 French Fast Cath
- Swan inserted for full Right Heart Cath (.025 wire used if needed to get to PA)
- Dr. Herr prefers Left PA for the device
- Pulmonary Arteriogram for anatomy and placement planning (Glowtape provides reference point)
- Open device, flush wire lumen, and whisk in saline for 10-30 seconds
- Insert device and advance to PA
- Device is deployed, delivery system removed, Steelcore pulled back to common PA
- Swan advanced over steelcore to right PA for calibration numbers
- Cardiomems calibrated
- Sheath pulled manual pressure only.