Cardiomems: Difference between revisions
From Bay Area Structural Heart Wiki
No edit summary |
(→Open) |
||
Line 24: | Line 24: | ||
*~20cc contrast | *~20cc contrast | ||
*Glowtape (Usually behind room 6) | *Glowtape (Usually behind room 6) | ||
{{Warn| The glowtape expired, so it has been removed from the sterile packaging, and lives on the shelf in Room 6 above the CardioMems devices}} | |||
===Standby=== | ===Standby=== |
Revision as of 15:05, 16 March 2021
The Cardiomems device is used to passively monitor Pulmonary Artery pressures.
Procedure type in McKesson: (33289) IMPL WRLS PA PRESSURE SENSOR
- Under Diagnostic heart cath.
Reps
- Dan Kivelson 925 787-5551
- Robert Tubbs 669 232-4527
Dr. Herr
Open
- US probe cover
- Micropuncture
- 75cm or 180 cm Amplatz Super Stiff
- 7fr. And 9fr. dilators
- 12fr. FastCath sheath
- 7fr. TD swan
- .025 wire
- .018 x 300 Steelcore
- 2-port Pressure manifold
- ~20cc contrast
- Glowtape (Usually behind room 6)
Warning: The glowtape expired, so it has been removed from the sterile packaging, and lives on the shelf in Room 6 above the CardioMems devices
Standby
- Cardiomems Device
Procedure
- No need for contrast manifold, Dr. Herr prefers to inject with a syringe
- 75 Amplatz is in IR
- Place Glowtape under patient, slightly left of spine, mid chest
- No need to open package. Place under bedding.
- Right groin access with micropuncture and ultrasound
- Amplatz wire inserted in micropuncture sheath
- 7 and 9 French dilators
- 12 French Fast Cath
- Swan inserted for full Right Heart Cath (.025 wire used if needed to get to PA)
- Dr. Herr prefers Left PA for the device
- Pulmonary Arteriogram for anatomy and placement planning (Glowtape provides reference point)
- Open device, flush wire lumen, and whisk in saline for 10-30 seconds
- Insert device and advance to PA
- Device is deployed, delivery system removed, Steelcore pulled back to common PA
- Swan advanced over steelcore to right PA for calibration numbers
- Cardiomems calibrated
- Sheath pulled manual pressure only.