Penumbra Thrombectomy: Difference between revisions

From Bay Area Structural Heart Wiki
(Created page with "This is the page for the Penumbra device. Coming Soon!")
 
 
(4 intermediate revisions by the same user not shown)
Line 1: Line 1:
This is the page for the Penumbra device.
Manufacturer: [[Has Manufacturer::Penumbra]]


Coming Soon!
Product website: [[Has Info Page::https://www.penumbrainc.com/peripheral-device/indigo-cat-rx/ | Penumbra Cat RX info page]]
 
Instruction sheet: [[Has Info Page::Internal or external direct link to setup pdf or similar]]
 
IFU: [[Has IFU:: https://www.penumbrainc.com/peripheral-device/indigo-cat-rx/ | Cat RX IFU]]
 
 
==Setup==
 
 
===Supplies===
*Penumbra engine(vacuum pump)
*Cannister
*Cat RX kit
 
<!-- You can upload images, and add them using the format below, otherwise delete this area.  -->
[[File:Penumbra.jpg|thumb|150px|right|Penumbra System|link={{filepath:Penumbra.jpg}}]]
 
===Circulator===
#Get Engine unit from behind room 1
#Bring a canister and the Cat RX box
#Drop kit contents on sterile field
#Plug unit in and turn on
#Connect passed off tubing to canister
 
===Scrub===
#Open package
#Pass off tubing to circulator
#Per rep: remove switch from tubing
#Attach catheter to tubing
#Open to vacuum when exiting guide, keep vacuum on until out of guide catheter outside the body
 
==Breakdown==
#Dispose of tubing and canister
#Clean engine and return to sub-sterile behind room 1
 
==General Notes==
*Switch is optional, it can cause clogs
*Pinch tubing if switch isn’t used before insertion to test vacuum
 
<br />
[[Category:Products]]

Latest revision as of 18:56, 12 January 2021

Manufacturer: Penumbra

Product website: Penumbra Cat RX info page

Instruction sheet: Internal or external direct link to setup pdf or similar

IFU: Cat RX IFU


Setup

Supplies

  • Penumbra engine(vacuum pump)
  • Cannister
  • Cat RX kit
Penumbra System

Circulator

  1. Get Engine unit from behind room 1
  2. Bring a canister and the Cat RX box
  3. Drop kit contents on sterile field
  4. Plug unit in and turn on
  5. Connect passed off tubing to canister

Scrub

  1. Open package
  2. Pass off tubing to circulator
  3. Per rep: remove switch from tubing
  4. Attach catheter to tubing
  5. Open to vacuum when exiting guide, keep vacuum on until out of guide catheter outside the body

Breakdown

  1. Dispose of tubing and canister
  2. Clean engine and return to sub-sterile behind room 1

General Notes

  • Switch is optional, it can cause clogs
  • Pinch tubing if switch isn’t used before insertion to test vacuum