AngioVac: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
Line 18: Line 18:
*AngioVac Cannula
*AngioVac Cannula
*Angiovac Circuit
*Angiovac Circuit
==Procedure==
#Access dependent on physician
#Amp SS wire and possible pre-close (Ask)
#Dilate
#Left groin usually return (Aortic) cannula
#*Connect aortic cannula to short quick-connect adaptor, then to return portion of circuit
#Right groin Dryseal
#Working wire placed, then advance AngioVac catheter.
#*Ensure good seal with Tuohy-borst adapter when entering Dry-seal
==Notes==
#Quick-connect adaptors on circuit can be attached to “flash” circuit to avoid clots in circuit while procedure paused
#Air on AngioVac side of circuit not a problem while getting started




[[Category:Procedures]]
[[Category:Procedures]]

Revision as of 16:09, 12 October 2020

AngioVac is a deep vein thrombosis removal system. Similar to ECMO, an external circuit is created to circulate blood through a filter. Clot is caught in the filter, and blood is returned to the body.

Access for the AngioVac cannula will generally be Right IJ or Right Femoral Vein. The return cannula will be the opposite leg vein generally.

Supplies

  • Amplatz Super Stiff 180 (7cm tip)
  • Micropuncture
  • (2) Perclose
  • Probe Cover
  • Indeflator
  • (2) 0-Ethibon
  • (2) Stopcock
  • 1 PDSII suture on a TP1 needle
  • 26fr. DrySeal
  • ECMO instruments
  • Arterial cannula Non-vented (Ask for size)
  • 20fr. Dilator
  • AngioVac Cannula
  • Angiovac Circuit

Procedure

  1. Access dependent on physician
  2. Amp SS wire and possible pre-close (Ask)
  3. Dilate
  4. Left groin usually return (Aortic) cannula
    • Connect aortic cannula to short quick-connect adaptor, then to return portion of circuit
  5. Right groin Dryseal
  6. Working wire placed, then advance AngioVac catheter.
    • Ensure good seal with Tuohy-borst adapter when entering Dry-seal

Notes

  1. Quick-connect adaptors on circuit can be attached to “flash” circuit to avoid clots in circuit while procedure paused
  2. Air on AngioVac side of circuit not a problem while getting started