Tricuspid ViV: Difference between revisions

From Bay Area Structural Heart Wiki
No edit summary
No edit summary
Line 15: Line 15:
|}
|}


== Supplies ==
==Supplies==


* Standard TAVR backtable setup
*Standard TAVR backtable setup
* 8F Sheath
*8F Sheath
* 150 J
*150 J
* 260J
*260J
* Safari
*Safari
* Alligator Clip Pacing cable
*Alligator Clip Pacing cable
* 6F Angled Pigtail
*6F Angled Pigtail
* 0 Prolene
*0 Prolene
* Micropuncture (for woggle only)
*Micropuncture (for woggle only)




Standby:  
Standby:  


* Agilis
*Agilis
* 6F balloon wedge
*6F balloon wedge


<br />
<br />


== Procedure ==
==Procedure==


* Single Venous access
*Single Venous access
* Single Preclose
*Single Preclose
* 8F Sheath inserted
*8F Sheath inserted
* Pigtail and J wire advanced to RV
*Pigtail and J wire advanced to RV
* J wire exchanged for Safari
*J wire exchanged for Safari
** Shave back of Safari for pacing
**Shave back of Safari for pacing
* Valve delivery system advanced
* Sheath exchanged for E Sheath
* Pacer tested
*Valve delivery system advanced
* Valve deployed
*Pacer tested
* Safari exchanged for J wire
*Valve deployed
* Perclose completed and woggle placed.
*Safari exchanged for J wire
* Voila.
*Perclose completed and woggle placed.
*Voila.


<br />
<br />

Revision as of 00:01, 12 September 2020

Anesthesia Imaging Access ACIST Manifold
MAC Fluoro, TTE

Femoral Vein

No No

Supplies

  • Standard TAVR backtable setup
  • 8F Sheath
  • 150 J
  • 260J
  • Safari
  • Alligator Clip Pacing cable
  • 6F Angled Pigtail
  • 0 Prolene
  • Micropuncture (for woggle only)


Standby:

  • Agilis
  • 6F balloon wedge


Procedure

  • Single Venous access
  • Single Preclose
  • 8F Sheath inserted
  • Pigtail and J wire advanced to RV
  • J wire exchanged for Safari
    • Shave back of Safari for pacing
  • Sheath exchanged for E Sheath
  • Valve delivery system advanced
  • Pacer tested
  • Valve deployed
  • Safari exchanged for J wire
  • Perclose completed and woggle placed.