==Procedure== | ==Procedure== | ||
#Plug console in and turn on. | #Plug console in and turn on. | ||
#Log into the device username: cpmcvn, password:cmpcvn | #Log into the device username:cpmcvn, password:cmpcvn | ||
#If device was powered off less than 5 minutes ago, it will ask if you want to continue the same patient. | #If device was powered off less than 5 minutes ago, it will ask if you want to continue the same patient. | ||
#Select the room you are in. | #Select the room you are in. | ||
Line 30: | Line 30: | ||
#Buttons at bottom of screen allow for switching between review and live screens | #Buttons at bottom of screen allow for switching between review and live screens | ||
#Unit is not set up for archive as of yet. Record results in McKesson | #Unit is not set up for archive as of yet. Record results in McKesson | ||
#Unplug unit to turn off. This is OK per the rep. | |||
===Troubleshooting=== | ===Troubleshooting=== | ||
#Turn everything off | #Turn everything off (Unplug) | ||
#Turn recording console back on, and re-pair with wireless transmitter | #Turn recording console back on, and re-pair with wireless transmitter | ||
[[Category:Setups]] | [[Category:Setups]] |
Revision as of 19:02, 10 June 2020
This is the iFR and FFR wire currently used in the cath lab.
- Unit will generally be kept behind Room 6
Procedure
- Plug console in and turn on.
- Log into the device username:cpmcvn, password:cmpcvn
- If device was powered off less than 5 minutes ago, it will ask if you want to continue the same patient.
- Select the room you are in.
- Enter patient info. (Can be entered at the end for STEMI)
- Follow on-screen prompts
- Re-zero PA pressure on device and (Optional)McKesson
- Flush wire and place flat on table
- Press connect on machine, then turn on wire transmitter
- Light should be solid green if connected
- Introduce wire into guide and advance until radio-lucent portion is outside the guide
- Equalize
- Select mode at top, center of screen
- RFR - Resting full-cycle Ratio (AKA - IFR)
- Average of 5-beat difference through full cardiac cycle
- Camera icon gives a single reading
- Pullback button activates pullback
- FFR - Traditional measurement using Adenosine
- Follow physicians guidance for Adenosine and recording
- Unit will display lowest measurement from recording
- RFR - Resting full-cycle Ratio (AKA - IFR)
- Bottom left of screen contains options for vessel and drug labels
- Buttons at bottom of screen allow for switching between review and live screens
- Unit is not set up for archive as of yet. Record results in McKesson
- Unplug unit to turn off. This is OK per the rep.
Troubleshooting
- Turn everything off (Unplug)
- Turn recording console back on, and re-pair with wireless transmitter