Quantien RFR wire (Abbott): Difference between revisions
From Bay Area Structural Heart Wiki
(Created page with "This is the iFR and FFR wire currently used in the cath lab. #Log into the device username: cpmcvn, password:cmpcvn #If device was powered off less than 5 minutes ago, it w...") |
No edit summary |
||
Line 1: | Line 1: | ||
This is the iFR and FFR wire currently used in the cath lab. | This is the iFR and FFR wire currently used in the cath lab. | ||
*Unit will generally be kept behind Room 6 | |||
==Procedure== | |||
#Plug console in and turn on. | |||
#Log into the device username: cpmcvn, password:cmpcvn | #Log into the device username: cpmcvn, password:cmpcvn | ||
#If device was powered off less than 5 minutes ago, it will ask if you want to continue the same patient. | #If device was powered off less than 5 minutes ago, it will ask if you want to continue the same patient. | ||
Line 18: | Line 21: | ||
#**Pullback button activates pullback | #**Pullback button activates pullback | ||
#*FFR - Traditional measurement using Adenosine | #*FFR - Traditional measurement using Adenosine | ||
#**Follow physicians guidance for Adenosine and recording | |||
#**Unit will display lowest measurement from recording | |||
#Bottom left of screen contains options for vessel and drug labels | |||
#Buttons at bottom of screen allow for switching between review and live screens | |||
#Unit is not set up for archive as of yet. Record results in McKesson | |||
===Troubleshooting=== | |||
#Turn everything off | |||
#Turn recording console back on, and re-pair with wireless transmitter |
Revision as of 21:05, 9 June 2020
This is the iFR and FFR wire currently used in the cath lab.
- Unit will generally be kept behind Room 6
Procedure
- Plug console in and turn on.
- Log into the device username: cpmcvn, password:cmpcvn
- If device was powered off less than 5 minutes ago, it will ask if you want to continue the same patient.
- Select the room you are in.
- Enter patient info. (Can be entered at the end for STEMI)
- Follow on-screen prompts
- Re-zero PA pressure on device and (Optional)McKesson
- Flush wire and place flat on table
- Press connect on machine, then turn on wire transmitter
- Light should be solid green if connected
- Introduce wire into guide and advance until radio-lucent portion is outside the guide
- Equalize
- Select mode at top, center of screen
- RFR - Resting full-cycle Ratio (AKA - IFR)
- Average of 5-beat difference through full cardiac cycle
- Camera icon gives a single reading
- Pullback button activates pullback
- FFR - Traditional measurement using Adenosine
- Follow physicians guidance for Adenosine and recording
- Unit will display lowest measurement from recording
- RFR - Resting full-cycle Ratio (AKA - IFR)
- Bottom left of screen contains options for vessel and drug labels
- Buttons at bottom of screen allow for switching between review and live screens
- Unit is not set up for archive as of yet. Record results in McKesson
Troubleshooting
- Turn everything off
- Turn recording console back on, and re-pair with wireless transmitter