This is the iFR and FFR wire currently used in the cath lab.
- Log into the device username: cpmcvn, password:cmpcvn
- If device was powered off less than 5 minutes ago, it will ask if you want to continue the same patient.
- Select the room you are in.
- Enter patient info. (Can be entered at the end for STEMI)
- Follow on-screen prompts
- Re-zero PA pressure on device and (Optional)McKesson
- Flush wire and place flat on table
- Press connect on machine, then turn on wire transmitter
- Light should be solid green if connected
- Introduce wire into guide and advance until radio-lucent portion is outside the guide
- Equalize
- Select mode at top, center of screen
- RFR - Resting full-cycle Ratio (AKA - IFR)
- Average of 5-beat difference through full cardiac cycle
- Camera icon gives a single reading
- Pullback button activates pullback
- FFR - Traditional measurement using Adenosine
- RFR - Resting full-cycle Ratio (AKA - IFR)