*** THIS IS A RESEARCH CASE ; EQUIPMENT WILL NOT SCAN, PLEASE KEEP BARCODES OF ALL NON-SCANNABLE EQUIPMENT FOR ROSALIE WITH A PATIENT LABEL ***

As of today 1/25/2025 we have done ONE of these procedures ( on 3/29/2024 ). This page is rough draft and will be updated along the way.

Room Set Up ; 2 Manifolds needed for pressure only. No Acist. IV Pole for back table similarly to MitraClip with 1L pressure bag. Will need the MitraClip plate and stool over the patient; distance from zyphoid unknown...Radial prepped for pigtail. Rampart, yes if the deployment angles allow.

Supplies

  • TAVI Pack
  • Micropuncture
  • 8Fr Pinnacle
  • 2 Perclose
  • 6fr radial sheath (used for the pigtail for pressure monitoring)
  • Long taper dilators 16fr, 24fr for access site dilation in preparation for Guide introduction
  • 24fr DrySeal
  • 6Fr Balloon wedge PA catheter
  • VersaCross transeptal system
  • 10mm Mustang for septostomy or a 14mm Armada
  • Knu-Knit, Dermabond, Lido with EPI
  • 0.035 x 260 J-tip wire
  • (2) 5fr Pigtail catheters or 1 double lumen pigtail catheter (Langston)
  • RV Pacing setup - There is no wire in the LV that could be used for pacing
  • Stand support plate (example: plate used for MitraClip) placed under patient’s legs
  • 2 transducer sets (RHC; Ao/LV monitoring)
  • Implant and Delivery System Preparation -Back table
  • (3) Sterile Bowls (1000ml or larger)
  • 4L heparinized saline
  • (4) Luer lock syringes (50ml or larger)
  • (1) Pickup
  • (1) Y-Tuohy Borst Sidearm Connector
  • (1) 3-way Stopcock
  • (1) Pressure bag
  • (3) 6ft Sterile Hemostatic lines
  • IV pole on wheels
  • Extra sterile gowns and gloves available for 4C team

Bail Out Equipment List

(List provided by Vendor)

  • ECMO
  • IABP
  • Impella CP
  • Blood transfusion ability
  • Pacemaker interrogator if applicable per patient (for rapid pacing)
  • Peripheral vascular balloons – over the wire, 0.035” wire compatible Septal Closure and PVL Instrument List
  • Septal Closure Devices available – 2 per size: • Gore Cardioform PFO: 20mm, 25mm, 30mm • Abbott Amplatzer Cribaform: 18mm, 25mm, 30mm, 35mm • Corresponding delivery systems to match the above devices
  • PVL plug devices available - (for example: Amplatzer ADO2 in these sizes: 4x6mm, 5x4mm, 5x6mm, 6x4mm, 6x6mm) Corresponding delivery systems to match the above devices (must be at least 80cm in length)
  • 110cm Cook Ansel or Shuttle in both 4 French and 5 French