Laser CVX-300

From Bay Area Structural Heart Wiki
Revision as of 17:36, 7 November 2024 by Suddemj (talk | contribs)

PHILIPS CVX-300 QUICK-START CHECKLIST

SET-UP

  • Plug in CVX-300 and Turn Power Switch ON System will Complete Self-Test (5 minute warm up) (An audible beep will sound indicating the laser system is ready)
  • Connect Foot Pedal

OPERATION

  • Insert Catheter and Calibrate
    • Depress the Calibrate button, aim the catheter and 1-2 inches away from the energy detector
    • Depress the foot switch until the system completes calibration approximately 5 seconds) -The "Calibration Complete" or Cal OK" lamp is displayed
  • Set the Fluence and Rate Per Physician's Instructions (i.e. 50/50)
  • Select the "Ready" Mode
    • Depress foot pedal per Physician's instructions until lesion(s) modification is complete.
  • Select "Standby" Mode when the laser i or is removed from body.

After Treatment is Complete

REPORTING

  • Record "Pulses Delivered" in log book
    • Depressing the "Pulses Delivered" button enables the total number of pulses during a procedure to be visible in the display window
  • Record "Treatment Time" in log book
    • Depressing the "Treatment Time" button enables the total lasing time of the procedure to be visible in the display window

TROUBLESHOOTING

  • Laser System Will Not Turn ON
    • Verify power cord connections
    • Verify interlock plug installed
    • Verify main circuit breaker is in ON position
  • Laser Buzzes When Plugged In
    • Release red emergency button
  • Fault Codes 1-4, "Service", "Service Laser Soon", "Power Error"
    • Shut machine OFF, wait 30 seconds, then Turn ON
    • Call PHILIPS Customer Service @ 800-231-0978 for assistance

SHUTDOWN/STORAGE

  • Turn the Laser System OFF
    • Disconnect and store the power cord
    • Disconnect and store the footswitch
    • Close the catheter connector door
    • Clean the energy detector's face with alcohol prep
    • Cover the laser system with the dust cover